Have you been thinking about buying a zero-point dry-well? Forget those ugly-looking units the competition makes. You can get a great looking and great performing zero-point dry-well from Hart Scientific.
The Hart 9101 has three test wells for inserting more than one probe at a time. All three wells are stable to ±0.005 °C. One well accommodates changeable inserts for varying probe diameters.
The Model 9101 takes advantage of the latest solid-state cooling technology rather than relying on older, less reliable sealed-water-cell devices. This eliminates the possibility that the sealed water cell will freeze and burst while transporting the unit to field locations. And our solid-state cooler is run by an adjustable electronic controller that can be recalibrated in your lab for convenient recertification. Simply place a certified standards thermometer in one of the wells and, if needed, tweak the 9101 controller until the standards thermometer reaches equilibrium at 0 °C.
Since the unit is completely self-contained and doesn’t require any user settings, you can run it on demand for instant access to an accurate, traceable zero point. Set it up with the reference junction of a thermocouple for high-accuracy thermocouple measurements.
Less costly than refrigerated baths, more accurate and less problematic than ice baths, and more durable and better looking than competitive units using sealed-water cells, the Hart 9101 Zero-Point Dry-Well is a great choice for any calibration lab!
Specifications |
|
Temperature Range |
0 °C (32 °F) |
Stability |
±0.005 °C |
Total Instrument Error |
±0.02 °C, typical; ±0.05 °C max. (18–25 °C ambient) |
Stabilization Time |
Approx. 30 minutes |
Temperature Coefficient |
±0.005 °C/ °C amb. |
Size (HxWxD) |
311 x 216 x 150 mm |
Power |
115 VAC (±10 %), 1 A or 230 VAC (±10 %), 0.5 A, specify, 50/60 Hz, 125 W |
Well Dimensions |
2 wells 6.4 mm dia. x 152 mm D (0.25 x 6 in), 1 well 7 mm dia. x 152 mm D (0.28 x 6 in). Includes one set of telescoping inserts to provide various smaller diameters |
Weight |
5.4 kg (12 lb.) |
NIST-Traceable Calibration |
Data at 0 °C |
Ordering Information |
|
9101 |
Zero-Point Dry-Well (includes one set of telescoping inserts to provide various smaller diameters) |
2130 |
Spare Well-Sizing Tube Set |
9325 |
Rugged Carrying Case |
Buďte prvý, kto napíše príspevok k tejto položke.