Fluke Bm. QA-ES - tester elektrochirurgických prístrojov

Značka: Fluke Biomedical

QA-ES Series II analyzes electrosurgical units quickly and accurately. A wide load-resistance range provides 128 user-selectable loads, including very low loads for testing many of today’s ESUs.
An accuracy of ± 2 % of reading down to 20 mA guarantees reliable high-frequency leakage results. With capability to run an automatic power- distribution test in as little as 1 minute, the QA-ES works fast so technicians save time.
An Ansur QA-ES software plug-in allows users to create and automatically run tests, capture data, and produce easy-to-read reports with a PC.

Dostupnosť Opýtajte sa
Možnosti doručenia
Kód: 1328
€1 €0,83 bez DPH
Fluke Bm. QA-ES - tester elektrochirurgických prístrojov
Značka: Fluke Biomedical
Fluke Bm. QA-ES - tester elektrochirurgických prístrojov
Ukončená výroba
Doprava zadarmo pre každý nákup nad 80€ s DPH vytvorený cez náš eshop s doručením na Slovensko.
Doprava zadarmo
pre každý nákup nad 80€ s DPH vytvorený cez náš eshop s doručením na Slovensko.
Široké možnosti platby na dobierku, kartou online alebo bankovým prevodom.
Široké možnosti
platby a doručenia pre vaše pohodlie. Prijímame platobné karty, Google Pay alebo Apple Pay.
Tovar skladom odosielame v deň objednávky (do 13:00)
Tovar skladom
odosielame v deň objednávky kuriérom SPS alebo službou Packeta (objednávky do 13:00).
Záručný, pozáručný servis teplomerov, multimetrov a iného metrologického vybavenia.
Servis a kalibrácia
teplomerov, multimetrov a iného metrologického vybavenia vrátane záručného a pozáručného servisu.
  • Automatic power distribution measurement, including power, current, peak-to-peak voltage closed load only), and crest factor
  • Oscilloscope output
  • High-frequency leakage measurements with accuracy of ± 2 % of reading
  • 128 internal user-selectable test loads from 10 Ω to 5200 Ω
  • Foot-switch output for triggering the ESU under test
  • Ansur QA-ES software plug-in for automated test protocols
  • Large display
  • RS-232 and Centronic-printer interface
Biomedical Test   >   Performance Analyzers   >   Electrosurgery   >  
QA-ES Electrosurgery Analyzer
Technical Specifications
Generator Output
Continuous Operation: Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor
Single Operation: Single measurement after the set delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor
Power Distribution: Automatic measurement of power, current, peak-to-peak voltage (closed load only), and crest factor through a user-selectable load range
RF Leakage current: Provides connections and load configurations to measure HF leakage from both grounded and isolated equipment
RECQM: Test the “return electrode control quality monitoring” using the QA-ES internal loads
Modes of Operation
  Manual or remote controlled (via Ansur)
Measurement
  True RMS value of applied waveform
RMS Bandwidth
Instrumentation Only: 30 Hz to 10 MHz (-3 dB)
With loads: 30 Hz to 2.5 MHz (-3 dB)
Low Frequency Filter
  100 Hz filter to avoid low-frequency disturbance or interference
Current
  20 mA to 2200 mA
Current Accuracy
  20 mA to 2200 mA ± 2 % of reading
Load Resistance
  10 Ω to 2500 Ω in step of 25 Ω (@ dc), 2500 Ω to 5200 Ω in step of 100 Ω (@ dc)
Additional Fixed Load
  200 Ω 400 W for 30 s; max 15 % duty cycle
Crest Factor
  The higher of the two peak voltage measurements is used for computation
Range
  1.4 to 16 (V peak/V RMS)
Foot-Switch Output
  The foot switch output can be used to trigger the electrosurgical unit
Peak-to-Peak Voltage
0 kV to 10 kV (closed load only) accuracy: ± 10 %
Note: Measurement is taken between the active and dispersive electrodes with closed load only.
Oscilloscope Output
  5 V/A uncalibrated, 100 mA RF current minimum input
Ansur QA-ES Plug-In Remote Control
  All functions and tests in QA-ES may be performed from the PC.
User-Programmable Test Sequences
  Allows unlimited numbers of test sequences with user-programmable templates and test limits. These tests include power distribution test, output test, HF leakage, and RECQM verification.
Storage and Recall
  Protocol formats and data may be stored, recalled, printed out, or transferred.
Temperature
Operating: 15 ºC to 35 ºC (59 ºF to 95 ºF)
Storage: 0 ºC to 50 ºC (32 ºF to 122 ºF)
Display
  LCD graphic display
Alphanumeric Format
  8 lines x 40 characters
Graphic Mode
  240 x 64 pixel matrix
Display Control
  5 F-keys, enter, cancel, control knob, and an encoder
Data Input/Outputs
  Parallel printer port and bidirectional RS232
Power
  115/230 VAC; 48 to 66 Hz, 35 VA
Housing
  Metal case
Dimensions (WxDxH):
  34.2 cm x 39.5 cm x 13.2 cm (13.5 in x 15.6 in x 5.2 in )
Weight
  9.8 kg (21.6 lb)