Specifications |
Power |
|
Two 9 V alkaline batteries (eight hours min continuous power). |
|
Optional battery eliminator |
|
Normal-Sinus-Rhythm Waveform |
ECG Reference: |
ECG amplitudes specified for Lead II (calibration), from the baseline to the peak of the R-Wave; other leads proportional |
Normal Sinus Rhythm: |
12-lead configuration with independent outputs referenced to right leg (RL); output to ten Universal ECG Jacks, color-coded to AHA and IEC Standards |
Low-Level Amplitude: |
0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV) |
Amplitude Accuracy: |
± 2 % of setting Lead II |
High-Level Output: |
(available on the BP3 connector) 0.2 V/mV ± 5 % of the ECG-amplitude setting |
ECG Rate: |
30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM (Power-on default: 80 BPM) |
Rate Accuracy: |
± 1 % of setting |
ECG-Waveform Selection: |
Adult (80 ms) or pediatric (40 ms) QRS duration (Power-on default: adult QR/P-R interval of 0.16 seconds) |
Superimposed Artifact: |
50 and 60 Hz, muscle, baseline wander, respiration |
ST-Segment Elevation/Depression: |
Adult mode only, -0.8 mV to +0.8 mV, in 0.1-mV steps; and additional steps, +0.05 mV and -0.05 mV (Power-on default: elevation 0 mV) |
|
Pacemaker Waveform |
Pacer-Pulse Amplitude: |
1, 2, 5, 10 mV, ± 10 % (Power-on default: 5 mV) |
Pacer-Pulse Width: |
0.1, 0.5, 1.0, 1.5, 2.0 ms, ± 5 % (Power-on default: 1.0 ms) |
Pacing Rate: |
75 BPM |
Waveforms: |
(Power-on default: atrial waveform) |
|
- Atrial |
|
- Asynchronous 75 BPM |
|
- Demand with frequent sinus beats |
|
- Demand with occasional sinus beats |
|
- AV sequential |
|
- Noncapture (one time) |
|
- Nonfunction |
|
Arrhythmias |
Baseline NSR: |
80 BPM |
PVC Focus: |
Left focus, standard timing (except where specified) |
Supraventricular Arrhythmias: |
(Power-on default: atrial fibrillation, coarse) |
|
- Atrial fibrillation, coarse |
|
- Atrial fibrillation, fine |
|
- Atrial flutter |
|
- Sinus arrhythmia |
|
- Missed beat (one time) |
|
- Atrial tachycardia |
|
- Paroxysmal atrial tachycardia (PAT) |
|
- Nodal rhythm |
|
- Supraventricular tachycardia |
Premature Arrhythmias (all one-time events): |
(Power-on default: premature atrial contraction) |
|
- Premature atrial contraction (PAC) |
|
- Premature nodal contraction (PNC) |
|
- PVC1 left ventricular |
|
- PVC1 left ventricular, R on T |
|
- PVC2 right ventricular |
|
- PVC2 right ventricular, early |
|
- PVC2 right ventricular, R on T |
|
- Multifocal PVCs |
Ventricular Arrhythmias: |
(Power-on default: PVCs 6/min) |
|
- PVCs 6/min (Power-on default) |
|
- PVCs 12/min |
|
- PVCs 24/min |
|
- Frequent multifocal PVCs |
|
- Bigeminy |
|
- Trigeminy |
|
- Multiple PVCs (one-time run of 2PVCs) |
|
- Multiple PVCs (one-time run of 5 PVCs) |
|
- Multiple PVCs (one-time run of 11 PVCs) |
|
- Ventricular tachycardia |
|
- Ventricular fibrillation, coarse |
|
- Ventricular fibrillation, fine |
|
- Asystole |
Conduction Defects: |
(Power-on default: first-degree heart block) |
|
- First-degree heart block |
|
- Second-degree heart block |
|
- Third-degree heart block |
|
- Right-bundle-branch block |
|
- Left-bundle-branch block |
|
ECG-Performance Testing |
Amplitude: |
0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV) |
Pulse Wave: |
30, 60 BPM, with 60 ms pulse width |
Square Wave: |
2, 0.125 Hz (Power-on default: 2 Hz) |
Triangle Wave: |
2, 2.5 Hz |
Sine Wave: |
0.5, 5, 10, 40, 50, 60, 100 Hz |
R-Wave-Detection Waveform: |
Haver-Triangle |
R-Wave Rate: |
30, 60, 80, 120, 200, and 250 BPM (Power-on default: 60 BPM) |
R-Wave Width: |
20 to 200 ms (10-ms steps); additional steps, 8, 10, and 12 ms (Power-on default: 10 ms) |
Rate Accuracy: |
± 1 % |
Amplitude Accuracy: |
± 2 %, Lead II (exception, ± 5 % for R-Waves ≤ 20 ms) |
|
Respiration |
Rate: |
0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM (Power-on default: 20 BrPM) |
Impedance Variations (∆Ω): |
0.2, 0.5, 1, or 3 Ω (Power-on default: ∆ 1 Ω ) |
Accuracy Delta: |
± 10 % |
Baseline: |
500, 1000, 1500, 2000 Ω, Leads I, II, III (Power-on default: 1000 Ω) |
Accuracy Baseline: |
± 5 % |
Respiration Lead: |
LA or LL (Power-on default: LA) |
Apnea Selection: |
12, 22, or 32 s (one-time events), or continuous (Apnea ON = respiration OFF) (Power-on default: 12 s apnea) |
|
Blood Pressure |
Input/Output Impedance: |
300 Ω ± 10 % |
Exciter-Input Range: |
2 to 16 V RMS |
Exciter-Input-Frequency Range: |
DC to 5000 Hz |
Transducer Sensitivity: |
5 or 40 µV/V/mmHg (Power-on default: 5 µV/V/mmHg) |
Pressure Accuracy: |
± (2 % of setting + 2 mmHg) |
Static Levels, P1: - |
10, 0, 80, 160, 240, 320, 400 mmHg (Power-on default: 0 mmHg) |
Static Levels, P2: - |
10, 0, 50, 100, 150, 200, 240 mmHg (Power-on default: 0 mmHg) |
Static Levels, P3: - |
5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg) |
Static Levels, P4: - |
5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg) |
Dynamic Waveforms, P1: |
|
- Arterial: |
120/80 |
- Radial artery: |
120/80 |
- Left ventricle: |
120/00 |
- Right ventricle: |
25/00 |
Dynamic Waveforms, P2: |
|
- Arterial: |
120/80 |
- Radial artery: |
120/80 |
- Left ventricle: |
120/00 |
- Right atrium (central venous or CVP): |
15/10 |
- Right ventricle: |
25/00 |
- Pulmonary artery: |
25/10 |
- Pulmonary-artery wedge: |
10/2 |
- Left atrium: |
14/4 |
Dynamic Waveforms, P3: |
|
- Arterial: |
120/80 |
- Radial artery: |
120/80 |
- Left ventricle: |
120/00 |
- Right atrium (central venous or CVP): |
15/10 |
- Right ventricle: |
25/00 |
- Pulmonary artery: |
25/10 |
- Pulmonary-artery wedge: |
10/2 |
- Left atrium: |
14/4 |
Dynamic Waveforms, P4 (Swan-Ganz Sequence): |
Right atrium (CVP); right ventricle RV); pulmonary artery (PA); pulmonary-artery wedge (PAW) |
Respiration Artifact: |
BP delta change from 3 to 16 mmHg |
BP Output: |
Mini DIN 7-Pin |
|
Temperature |
Temperature: |
0 °C (32 °F), 24 °C (75.2 °F), 37 °C (98.6 °F), and 40 °C (104 °F) (Power-on default: 0 °C / 32 °F) |
Accuracy: |
± 0.1 °C |
Compatibility: |
Yellow Springs, Inc. (YSI) Series 400 and 700 |
Output: |
mini DIN 7-pin |
|
Cardiac Output (optional) |
Catheter Type: |
Baxter Edwards, 93a-131-7f |
Calibration Coefficient: |
0.542 (0 °C injectate), 0.595 (24 °C injectate) |
Blood Temperature: |
37 °C (98.6 °F) ± 2 % |
Injectate Volume: |
10 cc |
Injectate Temperature: |
0 °C or 24 °C ± 2 % value (Power-on default: 0 °C injectate) |
Cardiac Output: |
2.5, 5, 10 lpm ± 5 % (Power-on default: 2.5 lpm) |
Faulty-Injection Curve: |
(Waveform for simulation available) |
Left-to-Right-Shunt Curve: |
(Waveform for simulation available) |
Calibrated Pulse: |
1.5 ° for 1 second (37 ° * 35.5 °) (Waveform for simulation available) |
Repeatability: |
± 1 % |
|
Fetal/Maternal ECG (optional) |
Maternal Heart Rate (Fixed): |
80 BPM |
Fetal Heart Rate (Selectable): |
60, 90, 120, 140, 150, 210, and 240 BPM (Power-on default: 120 BPM) |
Fetal Heart Rate (IUP): |
140 BPM at beginning, then varying with pressure |
Intrauterine-Pressure Waveforms: |
- Uniform acceleration (140-175-140 BPM; rate change lagging IUP contraction by 30 s) |
|
- Uniform deceleration (140-100-140 BPM; rate change lagging IUP contraction by 30 s) (Power-on default) |
|
- Early deceleration (140-100-140 BPM; no IUP lag time) |
|
- Late deceleration (140-100-140 BPM, starting at IUP peak) |
Wave Duration: |
90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and returning to 0, ± 4 mmHg (max) |
IUP Period: |
2, 3, or 5 minutes; and manual (Power-on default: manual) |
|
Computer Setup |
Port: |
Bidirectional (Data Communications Equipment) RS232 |
Baud Rate: |
9600 |
Parity: |
None |
Stop Bits: |
1 |
Data Bits: |
8 |
|
Battery Replacement |
|
Warning for low-battery condition (batteries to be replaced at this time) |
|
Temperature |
Storage: |
-25 to 50 ºC |
Operation: |
10 to 40 ºC |
|
Humidity |
|
80 % max relative humidity |
|
Dimensions |
|
7.5” L x 7.25” W x 2” H |
|
(19 cm L x 18.415 cm W x 5 cm H) |
|
Weight |
|
|