Specifications |
Power |
|
120/250 Vac, 50 W average, 100 W peak, 50/60 Hz |
|
Analog Outputs |
Cuff Pressure: |
0 to 499.95 mmHg FS ± 1.0 % of reading, 10 mV/mmHg |
Pulse Pressure: |
0.0 to 5.0 mmHg FS ± 1.0 of reading, 1.0 V/mmHg |
|
Digital Manometer |
Pressure Range: |
- 499.75 to 499.75 mmHg |
Measurement Parameters: |
instantaneous and peak |
|
Displayed Graphics |
|
Dynamic real-time NIBP cuff-pressure waveform programmed peripheral pulse and envelope waveforms |
|
Display |
|
Alphanumeric Graphic Display (LCD) |
Alphanumeric Mode: |
8 lines x 40 characters |
Graphics Mode: |
64 vertical x 240 horizontal dot matrix, backlight with viewing angle adjustment |
|
Digital Interfaces |
RS232C/Serial: |
bidirectional. Downloads cuff measurement data and controls test features with a compatible computer or via the medTester 5000C with the medCheck option. |
Parallel Printer: |
Centronics compatible |
Pulse Sync: |
0 V dc to 5 V dc (TTL) |
|
Cuff Mandrel |
Interlocking Plastic Blocks: |
four cuff circumferences, including 39.5 cm (large adult), 33 cm (adult), 26.6 cm (small adult) and 20 cm (child); maximum cuff : 15.25 cm. |
Truncated Plastic Cylinders: |
Three neonatal cuff circumferences, including 14 cm, 10 cm, and 7.6 cm; maximum cuff width of 7.6 cm. |
|
Pop-off Valve Testing |
|
Automatic test for operation of the monitor’s relief valve |
Measurement Parameters: |
instantaneous and peak pressure |
Maximum Pressure: |
499.75 mmHg |
|
System Leak Testing |
Start Pressure: |
499.75 mmHg max |
Elapsed Time: |
60 s (fixed) |
Leak-Rate Range: |
0.25 mmHg/min to 499.75 mmHg/min |
Pump: |
2.0 liters/minute minimum (free flow) |
|
Accuracy |
Systolic/Diastolic Mean Arterial Pressure (MAP): |
1 % of target value |
Cuff Pressure: |
± 1 % of reading ± 1 mmHg |
Input Overpressure Limit: |
± 1500 mmHg |
|
Autosequences |
|
Up to five user-programmable sequences to test NIBP monitors with a specific series of CuffLink performance tests, including static pressure test, leak test, and pop-off test |
|
Up to eight adult-neonatal-arrhythmia dynamic NIBP selections, each of which can be cycled up to 99 times during the sequence. |
|
Printable test report. |
|
Displayed Real-Time Parameters |
Instantaneous Cuff Pressure: |
0 mmHg to 300 mmHg |
Peak Cuff Pressure: |
0 mmHg to 500 mmHg peak |
Inflate/Deflate Time: |
0.1 s to 999.9 s |
Inflate/Deflate Rate: |
0.1 mmHg/s to 999.9 mmHg/s |
Total Measurement Time: |
0 s to 999.9 s max |
Selected Heart Rate: |
selected systolic/diastolic (mean) target values |
|
User-programmed vertical and horizontal shifts |
|
Dynamic Noninvasive Blood Pressure |
|
Simulation of a range of normal, hypertensive and hypotensive dynamic noninvasive blood pressures for typical adult, infant, and neonatal patients. Generation of normal, bradycardia, and tachycardia rhythm selections with a wide range of user-programmable peripheral pulse amplitudes (weak, normal and strong). Compatible with oscillometric NIBP devices. |
Preprogrammed Target Value Selections: |
Adult Systolic/Diastolic (MAP) (mmHg): 60/30 (40), 80/50 (62), 400/65 (75), 120/80 (90), 150/100 (115), 200/150 (165) and 255/195 (215) |
Neonatal/Pediatric Systolic/Diastolic: |
- Above selections, excluding 255/195 and 200/150 |
Repeatability: |
± 1 % of selected target value |
Adult Arrhythmia Selections: |
Baseline NIBP Target Value: (120/80) (NSR) |
|
Atrial fibrillation (A-Fib) |
|
Premature atrial contraction (PAC) |
|
Premature ventricular contraction (PVC) |
|
Missed beat (MB) |
|
Aberrant sinus conduction (AS) |
Preprogrammed Peripheral Pulse Waveforms: |
Pulse amplitude at MAP: 2.0 mmHg (typical adult value) |
|
Pulse volume range: 0.0 to 5.1 ml |
|
Pulse rise time: 80 ms (min) |
|
Heart rates (adult & neonate): 30, 40, 60, 80, 120, 160, 200 & 240 BPM |
|
Heart-rate accuracy: ± 1.0 % of selected rate |
User-Programmable Target Value Shifts: |
Horizontal Axis: preprogrammed target value selections shifted in 1.0 mmHg steps over a maximum range of ± 100 mmHg to increase or decrease dynamic pressure values |
|
Vertical Axis: relative amplitude shifted in 1 % increments over a maximum range from 0 to 200 % to simulate weak, normal, and strong peripheral pulses |
|
Dimensions (WxDxH): |
|
31.75 cm x 38.1 cm x 12.7 cm |
|
(12.5 in x 15 in x 5 in) |
|
Weight |
|