Specifications |
Display |
|
Bright, large 4-line x 40 character alphanumeric display with backlighting |
|
Pressure Generation/Measurement |
Static-pressure range |
0 mmHg to 400 mmHg(53 kPa) |
Difference Between Target Pressure and Actual Pressure: |
-5 mmHg |
Internal Leak Rate: |
< 2 mmHg per minute with minimum volume of 300 cc |
|
Four Respiratory Artifacts |
|
3 spontaneous breathing |
|
controlled ventilation |
|
3 Adult Wrist-Cuff Simulations |
|
Pressure Source |
|
Specified pressure generated from 0 mmHg to 400 mmHg in selectable increments of 1 mmHg |
|
Pressure Gauge |
|
Static pressure measured from 0 mmHg to 400 mmHg at the pressure port |
|
Pressure Relief Rest |
|
Test for the NIBPM pressure relief valve (0 mmHg to 400 mmHg) with display of peak pressure |
|
Neonate Internal Cuff Simulations |
|
Internal neonate cuff |
|
Four standard neonate pressures |
|
Normal Sinus Rhythm |
BP and ECG |
- Healthy heart |
|
- Weak pulse |
|
- Mild exercise |
|
- Strenuous exercise |
|
- Obese subject |
|
- Geriatric subject |
|
- Tachycardia |
|
- Bradycardia |
|
Irregular Pulse |
BP and ECG |
- Premature atrial contractions # 1 |
|
- Premature atrial contractions # 2 |
|
- Premature ventricular contractions |
|
- Atrial fibrillation and PVCs |
|
User-Definable Simulations |
|
User-defnable systolic and diastolic values, along with heart rate and pulse volume |
Ranges: |
|
- Systolic Pressure Range: |
20 to 250 |
- Diastolic Pressure Range: |
10 to 200 |
- Heart Rate: |
30 to 250 |
- Pulse Volume: |
0.1 cc to 2.4 cc |
|
Simulation Parameters Performance |
Max Pulse Volume: |
2.4 cc |
Max Heart Rate: |
- 200 BPM at 2.4 cc pulse volume |
|
- 250 BPM at 1.2 cc pulse volume |
Internal Neonatal Cuff Volume: |
20 cc |
Internal Adult Cuff Volume (Including NN Volume): |
310 cc |
Heart Rate Setting Accuracy: |
± 1 BPM |
Simulation Units: |
kPa and mmHg (user selectable) |
|
Pressure Leak Test |
|
The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the pressure loss over time. Peak pressure and present pressure are displayed at all times; leak rate is displayed when it is available. |
|
Autosequences |
|
Nine autosequences are provided for four tests and up to five simulations. |
|
Electrical ECG |
Signals: |
RA, LA, RL, LL, V |
Waveform: |
Lead II |
Amplitude: |
1 mV peak (± 10 %) |
Connections: |
Optional external ECG adapter |
|
Physiological synchronization with NIBP |
|
Serial Port |
|
Bidirectional RS232 port; baud rate of 9600 with no parity, one stop bit, and eight data bits. |
|
Pressure Measurement |
Pressure-Measurement Units: |
kPa, mmHg, cmH 2 0, cmH 2 O and psi (user selectable) |
Pressure-Measurement Range: |
0 mmHg to 400 mmHg |
Pressure-Measurement Resolution: |
0.1 kPa, 1.0 mmHg, 1.0 cmH 2 O, and 0.1 psi |
Pressure-Measurement Accuracy: |
|
- Standard Version (BP Pump 2 L ): |
- 0 to 300 mmHg: + 0.5 % of reading + 1 mmHg |
|
- 301 to 400 mmHg: + 2 % of reading |
- High-Accuracy Version (BP Pump 2 M ): |
- < 0.8 mmHg (0.1 kPa) throughout range |
|
Parallel Port |
|
25-pin female connector, with D-subminiature style and pinouts conforming to IBM "PC" printer port (unidirectional), HP and ASCII printers. |
|
Sample Adult Arm-Cuff Simulation Parameters (Standard) |
Standard Set of Blood Pressures |
|
- BP #1 |
Blood Pressure : 120/80 (93) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.68 cc |
- BP #2 |
Blood Pressure : 150/100 (116) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.65 cc |
- BP #3 |
Blood Pressure : 200/150 (166) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.60 cc |
- BP #4 |
Blood Pressure : 255/195 (215) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.55 cc |
- BP #5 |
Blood Pressure : 60/30 (40) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.75 cc |
- BP #6 |
Blood Pressure : 80/50 (60) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.70 cc |
- BP #7 |
Blood Pressure : 100/65 (76) |
|
Heart Rate : 80 |
|
Pulse Volume : 0.69 cc |
|
Dimensions |
|
10 in L x 10 in D x 5 in H |
|
(25.4 cm L x 25.4 cm D x 12.7 cm H) |
|
Weight |
|